Surfix Pro X

State-of-the-Art Measurement Technology – Dry Film Thickness Gauge with External Probe

Our new premium product meets the high demands of modern coating thickness gauges with numerous new features.

Surfix® Pro X offers easy handling in almost every application area, along with interchangeable probes and maximum precision.

Surfix Pro X is the new premium coating thickness gauge from Phynix. In addition to all the features of the Surfix Pro S model, this device offers a high-resolution color display, Bluetooth 4.0 and USB 2.0 interfaces, and a memory capacity of 50,000 measurements.

A wide variety of interchangeable probes provide multiple options for both universal applications and special measurement requirements. These probes, including one of the smallest measuring probes in the world, enable reliable measurements even in challenging geometries such as inside tubes, slits, and pipes.

F-type probes measure on ferrous metals like steel and iron using the magnetic induction method (TS 2311 / EN ISO 2178), while N-type probes are used for non-ferrous metals such as aluminum, bronze, and zinc using the eddy current method (TS 2674 / EN ISO 2360). FN-type probes combine both methods, allowing measurements on all types of metal substrates.

Another sector where the Surfix Pro X is used is the rubber and conveyor belt industry. With specially designed probes for measuring rubber thickness over steel mesh, it is ideal for determining the thickness of the rubber layers on the surface and sidewalls of tires.

Fields of Application
  • Electroplating
  • Paint shops
  • Automotive industry
  • Chemical industry
  • Aerospace industry
  • Shipbuilding industry
  • Rubber industry
  • Quality management and final inspections
  • Laboratories and research facilities
Advantages
  • High-resolution color display
  • Innovative and user-friendly measurements
  • Automatic recognition of base material
  • Automatic adjustment of the required measurement mode
  • Data storage for up to 100,000 readings and 1,000 individual calibrations with name, date, and time
  • Online statistics of all common parameters
  • Data transfer via USB 2.0 and Bluetooth 4.0
  • Intuitive user guidance – multilingual menus
  • Scan mode for continuous measurements
  • Optional plastic foot for up to 150°C or 300°C
  • Manufacturer inspection certificate
  • 2-year warranty
Product Code Description
11134 Phynix Surfix Pro X
Dry Film Thickness Gauge, probe not included
Range: 0–30,000 µm (depending on selected probe)
Resolution: 0.1 µm or < 0.2% of the reading (depending on selected probe)
Accuracy: ± 0.7 µm / 3 µm (depending on selected probe)
Calibration: Factory Calibration + Zero Calibration + Single Foil Calibration + Dual Foil Calibration + Offset Function (adding or subtracting a fixed value)
Memory: 50,000 measurements + statistical values such as average, standard deviation, minimum, and maximum
P10845 Prob FN 0,2
For coating thickness measurement on steel/iron (Fe) and non-ferrous (NFe) metals such as aluminum, bronze, and zinc
Range: 0–200 µm
Accuracy: ± 0.7 µm + 1% of the reading
P10749 Prob FN 1,5
For coating thickness measurement on steel/iron (Fe) and non-ferrous (NFe) metals such as aluminum, bronze, and zinc
Range: 0–1,500 µm
Accuracy: ± 1 µm + 1% of the reading
P10925 Prob FN 1,5 with acid resistant pole
For coating thickness measurement on steel/iron (Fe) and non-ferrous (NFe) metals such as aluminum, bronze, and zinc
Range: 0–1,500 µm
Accuracy: ±1 µm + 1% of the reading
P10949 Prob FN 1,5 R
For coating thickness measurement on steel/iron (Fe) and non-ferrous (NFe) metals such as aluminum, bronze, and zinc
Range: 0–1,500 µm
Accuracy: ±1 µm + 1% of the reading
P10751 Prob FN 1,5/90°
For coating thickness measurement on steel/iron (Fe) and non-ferrous (NFe) metals such as aluminum, bronze, and zinc
Range: 0–1,500 µm
Accuracy: ±1 µm + 1% of the reading
P10750 Prob FN 3,5
For coating thickness measurement on steel/iron (Fe) and non-ferrous (NFe) metals such as aluminum, bronze, and zinc
Range: 0–3,500 µm
Accuracy: ±2 µm + 1% of the reading
P10752 Prob F 1,5
For coating thickness measurement on steel/iron (Fe)
Range: 0–1,500 µm
Accuracy: ±1 µm + 1% of the reading
P10950 Prob F 1,5 R
For coating thickness measurement on steel/iron (Fe)
Range: 0–1,500 µm
Accuracy: ±1 µm + 1% of the reading
P10753 Prob F 3,5
For coating thickness measurement on steel/iron (Fe)
Range: 0–3,500 µm
Accuracy: ±2 µm + 1% of the reading
P10754 Prob F 10
For coating thickness measurement on steel/iron (Fe)
Range: 0–10,000 µm
Accuracy: ±3 µm + 1% of the reading
P10755 Prob N 1,5
For coating thickness measurement on non-ferrous (NFe) metals such as aluminum, bronze, and zinc
Range: 0–1,500 µm
Accuracy: ±1 µm + 1% of the reading
P10926 Prob N 1,5 with acid-resistant pole
For coating thickness measurement on non-ferrous (NFe) metals such as aluminum, bronze, and zinc
Range: 0–1,500 µm
Accuracy: ±1 µm + 1% of the reading
11002 Prob F30T
For tread and undertread rubber thickness
Range: Tread: 0–30 mm / Undertread: 2–10 mm
Accuracy: ±0.2 mm + 5% of the reading
11006 Prob F10-1
For measuring rubber thickness on a single steel belt layer (tire sidewall)
Range: 0–6.5 mm
Accuracy: ±0.1 mm + 3% of the reading
11007 Prob F10-2
For measuring rubber thickness on multi-layer steel belts (inside tire)
Range: 0–6.5 mm
Accuracy: ±0.1 mm + 3% of the reading
11008 Prob F10-3
Rubber thickness measurement on tire tread
Range: 0–6.5 mm
Accuracy: ±0.1 mm + 3% of the reading
11060 Prob F10-CP
Rubber thickness measurement including fabric carcass on tire sidewall, only for fabric carcass sidewall tires
Range: 0–10 mm
Accuracy: ±0.01 mm + 3% of the reading
11008 Prob F30-C
Rubber thickness measurement for steel cord conveyor belts
Range: 0–20 mm
Accuracy: ±0.2 mm + 3% of the reading